For Gas and Dust Atmospheres
These piezoresistive pressure transmitters are approved for use in highly explosion risk areas of group II (gas and dust).
The Y-line transmitters have an extremely small temperature error. This is achieved using an additional circuit containing a temperature sensor that subdivides the temperature range into fields that are 1,5 Kelvin (K) wide. The TK zero and TK compensation values are calculated for each field and programmed into the additional circuit. During operation, these values are fed into the analogue signal path depending on the temperature. Each temperature is the „calibration temperature” for this transmitter. The accuracy thereof is mainly determined by linearity. 120 fields are available, representing a possible temperature range of 180 K. The wider the temperature range, the greater the amount of testing that is needed to minimise the inaccuracy of the mathematical model.
General Data
| |
---|
量測範圍
|
0…100 mbar to 0…1000 bar
|
輸出信號
|
4...20 mA 0…5 V 0…10 V
|
精密度
|
± 0,25 %FS
|
儲存溫度
|
-40…100 °C
|
廠牌
|
Keller/瑞士製
|
Note:
The Series 23 (S) Ei, Series 25 (S) Ei, Series 26 Ei have been replaced by this Series.
The data sheet on the Series 23 (S) Ei, Series 25 (S) Ei, Series 26 Ei is yet available under:
Data Sheet Series 23 (S) Ei, Series 25 (S) Ei, Series 26 Ei